ACLS

Cardioversion vs. Defibrillation: Differences and Indications

Many physicians confuse synchronized cardioversion and defibrillation. This article explains indications, energy levels, device settings, and common pitfalls for both procedures.

Dr. med. univ. Daniel Pehböck, DESA

Author: Dr. med. univ. Daniel Pehböck, DESA

Specialist in Anesthesiology and Intensive Care Medicine, AHA-certified ACLS/PALS Instructor, Course Director Simulation Tirol

Reading time approx. 9 min

Electrical therapy for cardiac arrhythmias is among the most time-critical interventions in emergency medicine. Whether in the resuscitation bay, the intensive care unit, or the prehospital setting – the decision between synchronized cardioversion and defibrillation often has to be made within seconds. Yet clinical practice shows that confusing these two procedures is one of the most common and potentially most dangerous errors when using a defibrillator. An asynchronous shock in atrial fibrillation with a pulse can trigger ventricular fibrillation – a synchronized shock in ventricular fibrillation may not be delivered at all because the device cannot detect an R-wave trigger. This article systematically breaks down the differences and provides you with concrete decision-making tools for clinical practice.

The Underlying Physical Principle

Both cardioversion and defibrillation use the same mechanism: A transthoracic electrical shock simultaneously depolarizes a critical mass of myocardial cells. This interrupts the reentrant or chaotic electrical activity, giving the sinoatrial node the opportunity to resume its pacemaker function.

The crucial difference lies in the timing of energy delivery:

  • Defibrillation: The shock is delivered asynchronously – immediately after pressing the shock button, regardless of the point in the cardiac cycle.
  • Synchronized cardioversion: The device detects the R-wave on the ECG and delivers the shock synchronously – during the absolute refractory period of the ventricle.

This synchronization is not just a technical detail but a safety measure: If the electrical shock falls within the vulnerable phase (the relative refractory period, i.e., on or shortly after the T-wave), a so-called R-on-T phenomenon can trigger ventricular fibrillation. In a patient who initially has an organized rhythm with cardiac output, this would be an iatrogenic catastrophe.

Defibrillation: Indications and Procedure

When to Defibrillate?

Defibrillation is the treatment of choice for the two shockable rhythms in cardiac arrest:

  • Ventricular fibrillation (VF)
  • Pulseless ventricular tachycardia (pVT)

Both rhythms share the characteristic that there is no relevant cardiac output, and the electrical activity is either chaotic (VF) or organized but without ejection (pVT). In both cases, the goal is immediate, complete depolarization of the myocardium.

Why No Synchronization?

In ventricular fibrillation, there is no identifiable R-wave. The device in synchronization mode would search in vain for a trigger and not deliver the shock – a fatal loss of time. In pulseless ventricular tachycardia, an R-wave may be present, but the patient is clinically dead: There is no reason to wait for synchronization, and every second of delay worsens the outcome.

Energy Levels

The AHA guidelines recommend the following energy levels for defibrillation:

Device First Shock Subsequent Shocks
Biphasic 120–200 J (manufacturer-dependent) Same or escalating, up to maximum energy
Monophasic 360 J 360 J

In practice, biphasic devices dominate today. If you do not know the manufacturer-specific recommendation, choose 200 J as the standard energy for the first shock – this corresponds to the AHA recommendation when in doubt.

Sequence in the AHA Algorithm

  1. Rhythm analysis reveals VF/pVT
  2. Shock at the recommended energy (biphasic 120–200 J)
  3. Immediately resume CPR for 2 minutes – do not check for a pulse first
  4. After 2 minutes: repeat rhythm analysis
  5. If still VF/pVT: next shock, then CPR
  6. Epinephrine 1 mg IV/IO every 3–5 minutes (first dose after the second shock or as soon as IV/IO access is established)
  7. Amiodarone 300 mg IV/IO bolus after the third shock, with an optional additional 150 mg

Consistently minimizing interruptions to chest compressions – the so-called perishock pause – is essential. The goal is a perishock pause of less than 10 seconds.

Synchronized Cardioversion: Indications and Procedure

When to Cardiovert?

Synchronized cardioversion is used for tachyarrhythmias with a pulse when the patient is hemodynamically unstable. The AHA defines hemodynamic instability by the following signs:

  • Hypotension (systolic < 90 mmHg or clinically significant drop in blood pressure)
  • Acute altered mental status / loss of consciousness
  • Signs of acute heart failure (pulmonary edema)
  • Signs of ischemia (acute chest pain, ST changes)

The typical indication rhythms are:

  • Atrial fibrillation with rapid ventricular response and instability
  • Atrial flutter with instability
  • Supraventricular tachycardia (SVT), when vagal maneuvers and adenosine fail or the patient is unstable
  • Monomorphic ventricular tachycardia with a pulse, when unstable

Why Synchronization?

The patient has an organized rhythm with cardiac output. The shock is intended to terminate the rhythm without triggering ventricular fibrillation through unfortunate timing. Synchronization to the R-wave ensures that energy delivery occurs outside the vulnerable phase.

Energy Levels by Rhythm

The recommended energy levels vary depending on the arrhythmia and differ significantly from defibrillation:

Rhythm Biphasic Energy Notes
Narrow-complex tachycardia (SVT) 50–100 J initial Start low, escalate
Atrial fibrillation 120–200 J initial Higher energy needed than for flutter
Atrial flutter 50–100 J initial Often successful at low energy
Monomorphic VT with pulse 100 J initial Escalate if unsuccessful

If the initial shock fails, energy is increased stepwise. Based on current evidence, it is reasonable to escalate quickly to the maximum available energy after repeated failures.

Special Case: Polymorphic Ventricular Tachycardia

Polymorphic VT (e.g., Torsades de Pointes) occupies a special position: Although it is technically a VT with varying QRS morphology, it is treated like ventricular fibrillation – with asynchronous defibrillation at full energy. The reason is pragmatic: With constantly changing QRS morphology, the device cannot reliably identify the R-wave, and the rhythm frequently degenerates into ventricular fibrillation anyway.

Device Settings: What You Need to Watch For

Activating and Verifying Sync Mode

Most modern defibrillators start in asynchronous mode (defibrillation mode) by default. This is intentional because the most common use scenario – cardiac arrest – requires defibrillation. For cardioversion, you must actively enable the sync mode.

Key points for device handling:

  • Check the sync markers: After activating sync mode, the device displays markers on the monitor (usually arrows or dots) over the detected R-waves. Verify that these are correctly positioned on the R-waves and not on tall T-waves.
  • Adjust the lead and amplitude: If the R-wave amplitude is too low, the device will not detect it. Switch the lead or increase the amplitude.
  • Account for the delay: After pressing the shock button in sync mode, a few milliseconds pass until the next R-wave. Inform your team that the shock will be delayed – the shock button must be held down (on some devices).
  • Automatic reset: Many devices automatically reset the sync mode to asynchronous mode after each shock. If a second synchronized shock is needed, you must reactivate sync mode. This is one of the most common sources of error.

Paddles vs. Pads

Self-adhesive defibrillation electrodes (pads) are preferred over paddles because they ensure consistent electrode positioning, reduce the perishock pause, and allow continuous rhythm monitoring. This is particularly relevant for cardioversion because sync mode requires a stable ECG signal – motion artifacts from paddles can interfere with R-wave detection.

Common Pitfalls in Practice

Pitfall 1: Synchronized Cardioversion in Ventricular Fibrillation

As described above: The device searches for an R-wave, finds none, and does not deliver a shock. Valuable seconds are lost. Remember: Ventricular fibrillation = defibrillation = asynchronous = immediately.

Pitfall 2: Defibrillation in VT with a Pulse (Without Sync Mode)

You identify a monomorphic VT with a pulse, and the patient is unstable. You charge the device and press the shock button – without activating sync mode. The shock randomly falls on the T-wave, and the patient develops ventricular fibrillation. A treatable tachycardia has become a cardiac arrest.

Pitfall 3: Sync Mode Not Reactivated After the First Shock

The first synchronized shock does not convert the rhythm. You increase the energy and press the shock button again – but the device has automatically deactivated sync mode. The second shock is delivered asynchronously.

Pitfall 4: Incorrect R-Wave Detection

The sync markers are sitting on tall T-waves instead of R-waves. The shock is delivered exactly during the vulnerable phase – the opposite of the intended safety function. Always visually verify the markers.

Pitfall 5: Energy Too Low for Atrial Fibrillation

Atrial fibrillation generally requires higher energy levels than atrial flutter or SVT. An initial shock at 50 J for atrial fibrillation is frequently ineffective and leads to unnecessary repetitions.

Pitfall 6: Forgetting Sedation

Synchronized cardioversion is an elective procedure in a conscious patient (unless the patient already has an altered level of consciousness). Adequate procedural sedation – for example with propofol (0.5–1 mg/kg IV), midazolam (0.05–0.1 mg/kg IV), or etomidate (0.15–0.3 mg/kg IV) – is mandatory. Airway management must be ensured, and a provider capable of managing anesthesia should be present. In cardiac arrest, sedation is obviously not required.

Decision Algorithm: Defibrillation or Cardioversion?

The following algorithm summarizes the decision-making process:

  1. Does the patient have a pulse?

    • No → Cardiac arrest → Rhythm analysis
      • VF or pVT → Defibrillation (asynchronous, high energy)
      • Asystole or PEA → No shock indicated, CPR + epinephrine
    • Yes → Proceed to step 2
  2. Is a tachycardia present (HR > 150/min)?

    • No → Tachycardia unlikely to be the cause of instability → Search for other causes
    • Yes → Proceed to step 3
  3. Is the patient hemodynamically unstable?

    • No → Pharmacological therapy, expert consultation
    • Yes → Proceed to step 4
  4. What is the QRS morphology?

    • Regular, narrow → Synchronized cardioversion (50–100 J)
    • Irregular, narrow → Atrial fibrillation → Synchronized cardioversion (120–200 J)
    • Regular, wide → Monomorphic VT → Synchronized cardioversion (100 J)
    • Irregular, wide → Polymorphic VT → Defibrillation (asynchronous, full energy)

You should know this algorithm inside and out, because decision-making in an acute situation cannot take minutes.

Sedation and Team Communication

One aspect that is often overlooked in the heat of the moment is structured team communication. Before any electrical therapy, the following should be clearly verbalized:

  • Rhythm: "The patient has a monomorphic VT with a pulse."
  • Indication: "He is hemodynamically unstable – systolic blood pressure is 70."
  • Mode: "I am activating sync mode – please confirm that the markers are on the R-waves."
  • Energy: "I am charging 100 joules."
  • Safety: "Shock advisory – everyone clear from the patient!"

This closed-loop communication reduces errors and ensures that the entire team is on the same page.

Summary of Key Points

Defibrillation Synchronized Cardioversion
Mode Asynchronous Synchronous (R-wave-triggered)
Indication VF, pVT, polymorphic VT SVT, atrial fibrillation, atrial flutter, monomorphic VT with pulse (if unstable)
Patient No pulse (cardiac arrest) Pulse present, unstable
Energy High (120–200 J biphasic) Rhythm-dependent (50–200 J)
Sedation Not required Mandatory (if patient is conscious)
Timing Immediate Delayed (waits for R-wave)
Most common error Sync mode accidentally active Sync mode not activated or reset after shock

Practical Training

The distinction between cardioversion and defibrillation sounds straightforward in theory – but under the stress of a real emergency situation, with an unstable patient and a beeping monitor, correct execution becomes significantly harder. In the ACLS course from Simulation Tirol, you train exactly these scenarios in realistic simulations: rapid rhythm recognition, correct device settings, energy selection, and structured team communication. The courses follow the current AHA curriculum and give you the opportunity to make mistakes in a safe learning environment before they happen with a real patient.


Want to practice this hands-on?

In our ACLS-Kurs (Advanced Cardiac Life Support) you practice this topic hands-on with high-tech simulators and experienced instructors.

More Articles

In cooperation with

Netzwerk KindersimulationAmerican Heart Association · ERC Guidelines